Work

Communication Between Vascular Catheter Lumens: Safety Implications

Year: 2025
Type: article
Study design: Case Report
Abstract: ABSTRACT: In this case, we report a rare manufacturing defect in a multi-lumen central venous catheter, where communication between lumens was discovered during a pre-use flush check. This defect raises significant safety concerns, including potential medication interference and inaccurate CVP readings. Vigilant pre-use testing and heightened awareness of such anomalies are crucial to preventing life-threatening complications.
Language: English
Open Access status: diamond
OpenAlex ID: W4409485338
FWCI:
Cites: 4
Cited by:
Related to: 10