Communication Between Vascular Catheter Lumens: Safety Implications
Year: 2025
Type: article
Study design: Case Report
Abstract: ABSTRACT: In this case, we report a rare manufacturing defect in a multi-lumen central venous catheter, where communication between lumens was discovered during a pre-use flush check. This defect raises significant safety concerns, including potential medication interference and inaccurate CVP readings. Vigilant pre-use testing and heightened awareness of such anomalies are crucial to preventing life-threatening complications.
Source: Annals of Cardiac Anaesthesia
Authors: Gegal Pruthi, Ankita Dey, Dilraj Dhindsa
Institution: All India Institute of Medical Sciences
Language: English
Subfield: Emergency Medical Services
Field: Health Professions
Domain: Health Sciences
Open Access status: diamond
OpenAlex ID: W4409485338