Work

Detection limits for laser absorption spectroscopy of Li in laser ablation plumes

Year: 2025
Type: article
Abstract: A high-sensitivity laser absorption spectroscopy (LAS) method is demonstrated for rapid and non-contact analysis of Li in solids. Glass samples with Li concentrations ranging from 1 to 500 ppm by mass are ablated in 5 Torr air, and time-resolved Li atomic absorption is measured using a tunable laser near 671 nm. Significant improvements in analytical performance over prior laser-induced breakdown techniques are demonstrated. A single-shot limit of detection (LOD) of 180 ppb is shown for fixed-wavelength operation, improving to 6 ppb with averaging over 1000 ablation shots (100 s acquisition time). Isotope-resolved Li absorption spectra are measured with a 17 ppb LOD in 40 s. Methods for optimizing LOD based on noise/averaging properties of LAS experiments in LA plumes are discussed.
Language: English
Open Access status: green
OpenAlex ID: W4409569770
FWCI: 2.629
Cites: 25
Cited by: 8
Related to: 10