High Concordance of E-Nose-Derived Breathprints in a Healthy Population: A Cross-Sectional Observational Study
Year: 2025
Type: article
Study design: Observational Study
Abstract: Exhaled breath analysis using electronic noses (e-noses) is a promising non-invasive diagnostic tool. However, a lack of standardized protocols limits clinical implementation. This study evaluates the consistency of breathprints in healthy subjects using the Cyranose 320 e-nose to support standardization efforts. Breath samples from 139 healthy non-smoking subjects (age range 18-65 years) were collected using a standardized protocol. Participants exhaled into a Tedlar bag for immediate analysis with the Cyranose 320. Principal Component Analysis (PCA) was used to reduce data dimensionality, and K-means clustering grouped subjects based on breathprints. PCA identified four principal components explaining 97.15% of variance. K-means clustering revealed two clusters: 1 outlier and 138 subjects with highly similar breathprints. The median distance from the cluster center was 0.21 (IQR: 0.18-0.24), indicating low variability. Box plots confirmed breathprint consistency across subjects. The high consistency of breathprints in healthy subjects supports the feasibility of standardizing e-nose protocols. These findings highlight the potential of e-noses for clinical diagnostics, warranting further research in diverse populations and disease cohorts.
Source: Sensors
DOI: 10.3390/s25082610
Institution: University of Bari Aldo Moro
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $3,137
OpenAlex ID: W4409606600