Realistic approach for sub-bandage pressure appraisal
Year: 2025
Type: article
Abstract: /L0)*100 is the relative elongation expressed in [%]. Bandage specimens underwent a tensile test using a constant rate extension (CRE) device. The tensile test determined the elongation resulting from an arbitrary force and predicted the sub-bandage pressure. The elongation was then prescribed as the stretch required to achieve ex vivo pressure. The elongation and pressure were measured ex vivo with a ruler and the PicoPress® device, respectively, and compared to the predicted elongation and pressure. RESULT: The predicted pressure (24 mm Hg), calculated for ε = 48 % in the pressure equation, is not significantly different from the ex vivo pressure (26 mm Hg), as measured by a healthcare professional using the PicoPress®. The p-value (0.102) is greater than α (0.05) and the Pearson coefficient, R = 0.529, indicates a moderate relationship between the two variables.