Measurement of bilirubin in cerebrospinal fluid using the oxidase method on automated chemistry system advia XPT
Year: 2025
Type: article
Abstract: Background and aim: Evaluate the diagnostic performance of automated, quantitative bilirubin measurement, modified to extend its lower measurement ranges, in cerebrospinal fluid (CSF) using the Siemens analyzer Advia XPT. Results were compared with the gold standard spectrophotometry for diagnosis of subarachnoid haemorrhage (SAH). Method: Eighty clinical samples were analyzed on an Advia XPT, and results were compared to spectrophotometric results using the Agilent Cary 100 bio system. Method performance at low concentrations were evaluated using diluted control material and patient plasma and CSF samples. ROC curve analysis determined a suitable cutoff. Result: of 0.975), describing the relationship between measured and expected concentrations of diluted samples. The coefficient of variation, (CV), was 2.92 % at 0.598 μmol/L and 26.6 % at 0.161 μmol/L. Using the outcome of the analysis on Agilent Cary 100 as reference, sensitivity was 100 % and specificity 96 %, employing a cutoff of 0.41 μmol/L. Conclusion: Quantitative measurement of bilirubin in CSF using the bilirubin oxidase method on the automated Advia XPT platform perform well, with the analysis of low concentrations of bilirubin displaying a high precision and a high concordance with the results of spectrophotometry. These preliminary findings are indicative of the merits of quantitative measurement, that warrants further study of its diagnostic potential as an alternative to the more cumbersome spectrophotometry for diagnosing SAH.
Source: Practical Laboratory Medicine
Authors: Ida Branzell, Gabriella Lillsunde Larsson, Dieter Samyn, Paul Pettersson-Pablo
Institutions: Örebro University Hospital, Örebro University
Language: English
Subfield: Physiology
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $2,140
Funder: Universitetssjukhuset Örebro
OpenAlex ID: W4409798239