Year: 2025
Type: article
Abstract: Portable blood glucose meters (PBGMs) are essential for detecting glycemic disorders. In Brazil, veterinary devices are scarce, and human devices are used as an alternative. This study evaluated the accuracy of three human glucometers: Accu-Chek Guide® (ACG) and Accu-Chek Guide Me® (ACGM) (Roche), and EcoCheck® (EC) (Eco Diagnóstica). A total of 419 glycemic samples were collected and compared to the glucose oxidase method (GOM), considered the gold standard. Despite a strong positive correlation compared to the GOM (ACG r = 0.96; ACGM r = 0.9; EC r = 0.89; P <0.0001), when evaluated by the criteria defined by ISO15197:2013 for analytical precision, no PG obtained 99% of the readings within the range of +/- 15 for glucose levels <100 mg/dL and +/- 15% for glucose levels > 100mg/dL (ACG = 57.9%, ACGM = 47.9%, and EC = 28.4%). ACG was the only PG that fully met the clinical precision criteria according to ISO15197:2013, with 100% of its measurements in zones A+B of the error grid. Despite analytical inaccuracy, ACG PBGM proved to be the best among the devices tested in this study for use in dogs with whole blood samples, not generating any negative impact on animal health.
Source: Veterinary Sciences
Institution: Universidade Federal do Rio Grande do Sul
Language: English
Subfield: Physiology
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $2,534
OpenAlex ID: W4410242383