Work

Magnetic interference during biventricular pacemaker upgrade: A rare case of magnetized surgical instrument triggering automatic magnet response

Year: 2025
Type: article
Study design: Case Report
Abstract: Cardiac electronic implantable devices (CIEDs) utilize automatic magnetic response, a feature that allows for immediate and simple device reprogramming when exposed to a magnetic field. Placing a magnet over a CIED reprograms pacemakers to an asynchronous pacing mode and turns off tachycardia detections in implantable cardioverters-defibrillators (ICDs). This reprogramming is often strategically employed peri-operatively when electrocautery close to the leads is anticipated, as electric current delivered close to the device may inadvertently cause inappropriate pacemaker inhibition or inappropriate delivery of tachycardia therapies in ICDs due to electromagnetic interference (EMI).
Authors: Nitin Kondamudi, Jordan Kuester, Jordan M. Prutkin
Language: English
Field: Medicine
Open Access status: gold
Estimated APC paid: $1,100
OpenAlex ID: W4410735747
FWCI:
Cites: 6
Cited by:
Related to: 10