Year: 2025
Type: article
Abstract: The procedure for measurement uncertainty evaluation at humidity measuring instruments is considered. An expression for the deviations of hygrometer readings from the values set by the reference humid gas generator is written. Input quantities and their standard uncertainties are estimated. The calculation of the combined standard and expanded uncertainties when calibrating a hygrometer is performed. A budget for measurement uncertainty is compiled.
Source: Metrology and instruments
Author: Oleg Novoselov
Language: English
Subfield: Statistics, Probability and Uncertainty
Field: Decision Sciences
Domain: Social Sciences
Open Access status: diamond