Baseline values for Quantra® QPlus® in healthy pregnant women at term and comparison to standard laboratory coagulation values: a prospective observational study
Year: 2025
Type: article
Study design: Observational Study
Abstract: BACKGROUND: System, a novel point-of-care viscoelastic testing (POCVT) device, has been designed to provide rapid assessment of hemostasis. We aimed to evaluate the correlation between Quantra parameters and standard laboratory coagulation tests, and to establish baseline Quantra values in healthy term pregnant women. METHODS: Healthy pregnant women in labor or scheduled for elective cesarean delivery (CD) were enrolled in our prospective observational study. Blood samples for Quantra and standard laboratory coagulation tests were taken simultaneously. Quantra values, standard laboratory coagulation test, time of blood collected, and time to the result were recorded. We compared the baseline values between CD and labor group using a t-test, and the correlation between Quantra and standard laboratory coagulation test was calculated using partial Pearson correlation. RESULTS: 170 healthy pregnant women were included; 126 cases were in the CD group, and 44 patients were in the labor group. We found a strong correlation between Quantra Fibrinogen contribution to Clot Stiffness (FCS) and fibrinogen level (r = 0.67). The median [interquartile range] time of fibrinogen results by Quantra was 36 [28, 48] minutes faster than the standard laboratory coagulation tests. Baseline ranges for Quantra values, which were not significantly different between the two groups, demonstrated hyperfibrinogenemia during pregnancy. CONCLUSION: Quantra is a novel POCVT device that rapidly provides coagulation status in pregnant women. The strong correlation between FCS and fibrinogen level can be helpful for early recognition of hypofibrinogenemia for the management of PPH.
Source: BMC Pregnancy and Childbirth
Language: English
Subfield: Pulmonary and Respiratory Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $3,290
Funder: HemoSonics
OpenAlex ID: W4413055158