Active middle ear implants—Enhanced efficiency of oval window stimulation with larger piston size
Year: 2025
Type: article
Abstract: INTRODUCTION: Direct piston-stimulation of the oval window with a middle ear implant, following a stapedotomy or stapedectomy, is one treatment option for patients with otosclerosis. Here, we experimentally investigated whether increasing the surface area of the actuated piston enhances transmission efficiency in oval window (OW) stimulation. METHODS: ). Following stapedectomy, a piezoelectric transducer was used for OWC actuation. Additionally, a numerical model was also developed to estimate the output of the floating mass transducer (FMT, MEDEL, Austria). RESULTS: Across frequencies, the average pressure difference was 18 and 8 dB higher for the large and medium OWC compared to the small OWC, respectively. Statistical analyses showed significant differences (p < 0.05) between piston diameters across most frequencies. Numerical simulation results using the volume displacement at the OW showed smaller output differences from 0.3 to 10 kHz, indicating that leakage at open gaps after stapedectomy has to be taken into account. CONCLUSION: Our findings confirm that, for active stimulation with an actuator, increasing the surface area of the OWC enhances output efficiency proportionally. Nevertheless, volume leakage at open gaps to the OW may contribute additionally.
Source: Hearing Research
Language: English
Topic: Ear Surgery and Otitis Media
Subfield: Otorhinolaryngology
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $4,130
Awards: G2270855615, EXC 2177: Hearing4all
OpenAlex ID: W4413423518