Feasibility and limitations of a postal dosimetry audit system using RPLGD for high-dose-rate brachytherapy

Year: 2025
Type: article
Abstract: International recommendations and South Korean regulations emphasize the need for dosimetry audit systems to ensure treatment accuracy in high-dose-rate brachytherapy. In this study, we designed and tested a prototype polycarbonate postal audit phantom to assess its feasibility. Using six GD-302M radiophotoluminescent glass dosimeters. Dose measurements were performed at points A and B based on the treatment planning system (TPS) and PHITS. These were validated in the Korea laboratory accreditation scheme and secondary standard dosimetry laboratory using Co-60 beam. Reproducibility and linearity tests showed stable responses; however, a maximum deviation of 13.5 % was observed between the measured and TPS doses at point A. The PHITS simulations revealed that the TPS overestimated the dose by approximately 10 %, whereas glass dosimeters showed up to 20 % overestimation because of low-energy gamma rays from the Ir-192 source. Additional water phantom measurements demonstrated that the RPLGD can provide reasonably consistent dose readings under water-equivalent conditions. The discrepancies observed in the audit phantom were found to follow a consistent trend and could be assessed using normalization. The combined expanded uncertainties were calculated as 3.1 % and 3.2 % ( k = 2 ) for the source holder models 70010 and 72280, respectively. The audit phantom showed potential for application in nationwide postal audits, pending further validation. To ensure clinical applicability, future work should focus on addressing the energy dependence of the RPLGD and conducting a multi-institutional pilot study. • Many studies recommend audit systems to verify treatment accuracy through accredited organizations. • Lack of regulatory standards for dosimetry audits in South Korea may lead to safety risks. • A prototype postal dosimetry audit phantom was developed and tested to assess feasibility for future QA systems. • Uncertainty assessment was performed to evaluate the potential applicability of the system for postal audit purposes.
Language: English
Subfield: Radiation
Open Access status: hybrid
Estimated APC paid: $3,530
OpenAlex ID: W4413890275
FWCI:
Cites: 14
Cited by:
Related to: 10