Work

Prospective comparison of ultra-high-resolution photon-counting detector CT coronary angiography versus conventional CT

Year: 2025
Type: article
Abstract: OBJECTIVES: Determine differences in quantitative stenosis severity measurements for ultra-high-resolution (UHR) photon-counting detector (PCD) coronary CT angiography (cCTA) relative to energy-integrating detector (EID) cCTA in a large patient cohort. METHODS: Adult participants seen between November 2022 and March 2023 underwent a clinical dual-source EID-CT cCTA and a research dual-source PCD-CT cCTA on the same day. Percent diameter stenosis (PDS) was measured, and stenosis severity was assigned based on the PDS of the most severe lesion per patient to determine a coronary artery disease reporting and data system (CAD-RADS) score. Agreement between EID-CT and PCD-CT for PDS and CAD-RADS was determined. RESULTS: A total of 112 participants were enrolled, yielding 82 subjects with at least 1 stenosis at PCD-CT (mean age, 68.0 ± 10.8 years; 69.5% [57/82] males). A total of 177 paired stenoses were quantified. The percent decrease in mean PDS from PCD-CT (24.3%) to EID-CT (29.4%) was 17.3% (median difference in PDS: -5.0%, 95% CI: -5.7% to -4.0%). The assigned CAD-RADS score changed in 31/82 subjects. In 11/31 subjects, the most severe stenosis was missed at EID-CT due to partial volume averaging of a small calcification, yet PCD clearly identified stenosis at those locations. In 20/31 subjects, the CAD-RADS score decreased at PCD-CT due to the decreased calcium blooming resulting from the improved spatial resolution. CONCLUSIONS: UHR PCD cCTA decreases quantitative measures of stenosis severity when a stenosis is identified at EID, leading to decreases in CAD-RADS assignments. ADVANCES IN KNOWLEDGE: With its improved spatial resolution, PCD-CT identifies stenoses missed on EID-CT.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W4414014140
FWCI: 1.592
Cites: 32
Cited by: 3
Related to: 10