Work

Comparison of Two- and Three-Phase Devices Generating a Rotating Magnetic Fieldfor Magnetic Hyperthermia Applications

Year: 2025
Type: article
Abstract: This article describes systems generating high-frequency rotating magnetic fields for magnetic hyperthermia treatments. It covers two- and three-phase device systems powered by rectangular signals. A passive bandpass filter tuned to a specific frequency (100 kHz) is placed between the magnetic circuits and the DC power source powering the device. The paper compares the electrical parameters of both solutions, including the supply voltage, magnetic field strength amplitude H, and magnetizing current IL as a function of the supply voltage (Udc). At a fixed supply voltage Udc, the magnetizing current IL and the rotating magnetic field strength amplitude H are approximately twice as large for the three-phase system as for the two-phase system. The relationships between the magnetizing currents IL and the magnetic field strength amplitude H as a function of the supply voltage Udc are linear.
Source: Magnetism
Language: English
Open Access status: gold
Estimated APC paid: $1,207
OpenAlex ID: W4414093614
FWCI: 0.235
Cites: 16
Cited by: 1
Related to: 10