Work

Improving Proteomic Dynamic Range with Multiple Accumulation Precursor Mass Spectrometry

Year: 2025
Type: article
Abstract: Orbitrap (OT)-based mass spectrometer platforms are a gold standard in high-resolution mass spectrometry, where their primary disadvantage is slower-scanning speed in comparison to time-of-flight or linear ion trap mass analyzers. In this study, we utilize long OT transients to extend the precursor dynamic range by modifying the selected ion monitoring method to multiplex several precursor m / z ranges into a single scan, which we call “multiple accumulation precursor mass spectrometry”. Our approach requires no software or hardware modifications and hides the additional ion accumulation steps during the time it takes to make other Orbitrap measurements, producing precursor spectra with nearly 2× dynamic range and essentially no consequences. We collected data using both data-dependent acquisition (DDA) and data-independent acquisition (DIA) methods to evaluate a range of approaches. With DDA, MAP-MS precursor quantification improves with higher quality measurements. At the same time, DIA detection is enhanced by up to 11% when combining precursor and tandem mass spectra for peptide detection.
Language: English
Subfield: Spectroscopy
Field: Chemistry
Open Access status: green
OpenAlex ID: W4414162085
FWCI: 0.987
Cites: 32
Cited by: 4