Double-Guide Catheter “Criss-Cross” Technique and Left Main Bifurcation Rotational Atherectomy
Year: 2025
Type: article
Study design: Case Report
Abstract: BACKGROUND: Rotational atherectomy of severely calcified left main (LM) bifurcation lesions can be hazardous, especially when maintaining a protective guidewire in one branch is mandatory. CASE SUMMARY: An 82-year-old man underwent successful percutaneous coronary intervention of a calcified LM bifurcation using a double-guide catheter "criss-cross" technique to reduce periprocedural ischemia and iatrogenic dissection risk associated with a single larger catheter. Rotational atherectomy was safely performed on a tight proximo-ostial circumflex lesion while retaining a protective guidewire in the dissected left anterior descending artery. DISCUSSION: Although protective microcatheters have been described, bench studies show consistent microcatheter damage, risk of nonorganic material embolization, and problematic guidewire interactions. Proper positioning of guide catheters modifies wire bias, improving guidewire separation and minimizing burr-wire interactions. Bench testing confirmed this advantage. TAKE-HOME MESSAGES: The novel double-guide catheter criss-cross technique potentially increases safety in complex calcified proximo-ostial LM lesions by reducing guidewire interactions. Modification of wire bias could extend to all proximo-ostial coronary lesions.
Source: JACC Case Reports
Language: English
Subfield: Surgery
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $850
OpenAlex ID: W4414978944