Work

Proton Irradiation on Hydrogenated Amorphous Silicon Flexible Devices

Year: 2025
Type: article
Abstract: Radiation damage tests in hydrogenated amorphous silicon (a-Si:H) flexible flux and dose-measuring devices have been performed with a 3 MeV proton beam, to evaluate combined displacement and total ionizing dose damage. The tested devices had two different configurations and thicknesses. The first device was a 2 μm thick n-i-p diode having a 5 mm x 5 mm area. The second device was a 5 μm thick charge selective contact detector having the same area. Both the devices were deposited on a flexible polyimide substrate and were irradiated up to the fluence of 1016neq/cm2. The response to different proton fluxes has been measured before irradiation and after irradiation at 1016neq/cm2for charge-selective contacts and n-i-p devices. The effect of annealing for partial performance recovery at 100°C for 12 hours was also studied and a final characterization on annealed devices was performed. This test is the first combined displacement and total ionizing dose test on flexible a-Si:H devices.
Language: English
Open Access status: green
OpenAlex ID: W4415003196
FWCI:
Cites:
Cited by:
Related to: