Year: 2025
Type: article
Abstract: BACKGROUND: Heart valve diseases remain a leading cause of death in industrialized nations. Polycarbonate urethane (PCU) is a promising material for heart valve prostheses due to its biocompatibility and low calcification tendency. However, the impact of processing methods on calcification remains unclear. METHODS: PCU patches were fabricated via hot pressing or solution casting. Both groups (n = 3 each), along with bovine pericardium patches as positive controls (n = 3), were incubated for 10 weeks in a custom in vitro calcification fluid. Calcification, cytocompatibility, and material properties were assessed using light and electron microscopy, infrared spectroscopy, and gel permeation chromatography (GPC). RESULTS: Calcification was observed in hot-pressed PCU and control patches but not in solution-cast PCU. Both PCU types showed comparable cytocompatibility. Spectroscopy and GPC revealed chemical and structural changes in hot-pressed PCU, likely promoting calcification. CONCLUSION: Hot pressing alters the chemical structure of PCU and increases its calcification propensity without affecting cytocompatibility. These findings highlight the importance of process control and in vitro screening during heart valve material development.
Source: Artificial Organs
DOI: 10.1111/aor.70028
Authors: Jan Ritter, Christoph Schmitz, Stephan Rütten, Abdelhafid Aqil, Cécile Oury, Thomas Schmitz‐Rode, Willi Jahnen‐Dechent, Ulrich Steinseifer, Johanna C. Clauser
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,990
OpenAlex ID: W4415147308