Fundamental Sensor Response Time Limitations of Practical Air Temperature Measurement

Year: 2025
Type: article
Abstract: Air temperature measurements in naturally ventilated thermometer screens underpin the instrumental climate record. Increasing automation is, however, revealing limitations. One is through thermometer time response, especially in light winds or calm conditions, often at the daily temperature minimum. The exponential time response for thermometers enclosed within a Stevenson screen is a key parameter, but poorly known. Here, is evaluated in a practical experimental situation against the World Meteorological Organization (WMO)'s recommended sensor s. We find increases with sensor diameter , with only a = 2 mm sensor meeting WMO expectations, even then requiring ambient wind speeds . Typical = 4 mm sensors never meet the criterion when either force‐ or naturally ventilated, with mins in a naturally ventilated arrangement under calm conditions. Inadequate will lead to underestimation of the diurnal temperature range or other local measures derived from daily temperature maxima and minima.
Language: English
Open Access status: gold
Estimated APC paid: $3,190
OpenAlex ID: W4415470280
FWCI: 1.309
Cites: 16
Cited by: 3