Year: 2025
Type: article
Abstract: Background. Serum-sodium reporting tolerates a total allowable error (TEa) of only ±4 mmol/L, yet many laboratories continue to operate at the marginal three-sigma level because the quantitative benefit of additional sigma capability is poorly characterized. Objectives. The study aims to translate sigma metrics into clinically intuitive risk estimates by (1) quantifying the proportion of QC results that exceed the TEa at five sigma levels (2 – 6 σ) and (2) determining whether successive sigma gains produce statistically significant reductions in error. Methodology. Five (5) hypothetical assays were parameterized with a common mean of 140 mmol/L and CVs corresponding to 2-, 3-, 4-, 5- and 6-sigma performance. For each assay, 1,000 Monte-Carlo iterations were run, each iteration simulating 36,500 QC results (assuming 100 runs/day for 365 days) drawn from N(µ = 140, σ = µ × CV). The error rate (the proportion of results outside ±4 mmol/L) was recorded per iteration. Distributions were summarized (mean, range, SD); differences were evaluated with one-way ANOVA followed by Tukey’s HSD. Results. Mean (±SD) error rates declined significantly with increasing sigma: Assay A (2σ): 0.0456 ± 0.0011; Assay B (3σ): 0.00270 ± 0.00027;Assay C (4σ): 6.3 × 10⁻⁵ ± 4.1 × 10⁻⁵; Assay D (5σ): 5.8 × 10⁻⁷ ± 8.0 × 10⁻⁷; and Assay E (6σ): 2.0 × 10⁻⁷ ± 3.1 × 10⁻⁷. The maximum single-iteration error rate fell from 0.0505 at 2 σ to 1.1 × 10⁻⁴ at 4 σ. The 5σ and 6σ processes produced zero TEa violations in ≥96 % of iterations. ANOVA confirmed a global difference (p < 0.001); all pairwise contrasts were significant (p < 0.001) except between 5 σ vs 6 σ (p = 0.62). Conclusions. Each one-sigma gain yields an order-of-magnitude reduction in TEa violations until a plateau is reached at ≥5 σ, where residual analytical risk is negligible. These simulations support the recommendation that laboratories operating serum-sodium assays below 4 σ should prioritize precision improvements or enhanced QC strategies, whereas ≥5 σ assays may safely adopt less intensive QC without compromising patient safety.
Source: Philippine Journal of Pathology
DOI: 10.21141/pjp.2025.08
Authors: Mark Angelo Ang, Karen Cybelle Sotalbo
Institution: University of the Philippines Manila
Language: English
Subfield: Physiology
Field: Medicine
Domain: Health Sciences
Open Access status: diamond
OpenAlex ID: W4415533183