Implementation, Validation and Clinical Testing of Oximetry Device for Microcirculation Assessment in Oral Tissue
Year: 2025
Type: article
Abstract: The recent rise in living standards has been accompanied by increased awareness and emphasis on oral health. Non-invasive assessment of gingival microcirculation and accurate evaluation of oxygen supply to oral tissues are critical for the early diagnosis of oral diseases. These factors also play a pivotal role in optimizing treatment planning and improving outcomes in dental implantology. In this study, we report the development and implementation of a novel pulse oximetry device based on reflective photoplethysmography technology, designed for non-invasive, real-time monitoring of gingival health through the measurement of oxygen saturation levels. A detailed description of the technology, including key aspects of sensor probe design, is provided, with particular emphasis on the calibration process and performance evaluation of the prototype. Furthermore, we present and discuss the first proof-of-concept gingival oxygen saturation measurements obtained in a clinical setting during oral rehabilitation consultations.
Source: Sensors
DOI: 10.3390/s25216604
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $3,137
OpenAlex ID: W4415650227