Year: 2025
Type: article
Abstract: Purpose: The maximal isometric squat assessment is frequently used by clinicians and coaches in both rehabilitation and performance settings. The most common technology used during this assessment is force plates; however, this comes with several limitations regarding portability, set-up and cost. The purpose of this study therefore was to determine the intra-and inter-session reliability of a commercially available portable fixed dynamometer during an isometric squat assessment. Methods: Ten male participants (age, 25.3 ± 4.6 y; body mass, 82.7 ± 11.9 kg; height, 177.8 ± 9.0 cm) completed an isometric squat test across three testing occasions, each separated by seven days. Peak force, mean force, impulse, and peak rate of force development (PRFD) were the variables of interest. Results: There was no significant (p < 0.050) systematic change within the variables across the three trials and sessions. Peak force, mean force and impulse were found to have good to excellent intra-(ICC = 0.87-0.98,CV = 3.1-8.7%)and inter-session reliability (ICC = 0.92-0.96,CV = 5.2-8.5%).PRFD was found to have good to excellent intra-and intersession relative consistency (ICC = 0.84-0.95);however, absolute consistency was not found acceptable (CV = 16.1-28.5%). Conclusion: It appears that a portable fixed dynamometer can be used to reliably assess peak force, mean force, and impulse during an isometric squat strength in a healthy male population; however, PRFD should be interpreted with caution.