Method Validation of Dipeptidyl Peptidase 3 Assay

Year: 2025
Type: article
Abstract: BACKGROUND: Dipeptidyl peptidase 3 (DPP3) is a peptidase released from dying cells. It cleaves proteins in the renin-angiotensin pathway, which can result in hemodynamic instability. At elevated concentrations DPP3 is associated with worse outcomes, particularly in patients with shock. Herein we describe the assay performance of a DPP3 assay (4TEEN4 Pharmaceuticals GmbH) in human plasma. METHODS: DPP3 concentration was measured using the DPP3 immunoluminometric assay (4TEEN4 Pharmaceuticals GmbH) and the signal was read using a luminometer (Berthold Centro LB963). Analytical performance was established for precision, linearity, accuracy, detection limit, analytical specificity, reference interval, kit lot-to-lot comparison, specimen type, and sample stability. RESULTS: Limit of detection was verified at 1.6 ng/mL in EDTA plasma with a coefficient of variation (CV) of <10%. Precision studies revealed a CV ≤ 6% at 28.5 ng/mL and 59.4 ng/mL and comparability with a manufacturer performed assay was demonstrated between 7.7 and 195.2 ng/mL. An upper 97.5% limit of 22 ng/mL without age or sex associations was verified in healthy donors. The assay was not susceptible to interference from lipemia or bilirubin. However, measured DPP3 concentrations increased linearly with increasing hemolysis. DPP3 concentrations are stable in EDTA plasma for up to 24 h and at least 11 months when stored ambient or at -80°C, respectively. CONCLUSIONS: DPP3 can be measured precisely in EDTA plasma using the immunoluminometric DPP3 assay. Given the potential clinical use of DPP3 in critical care patients, caution should be taken to avoid inducing pre-analytical hemolysis during sample collection.
Institution: Mayo Clinic
Language: English
Subfield: Physiology
Field: Medicine
Open Access status: hybrid
Estimated APC paid: $4,212
OpenAlex ID: W4416186727
FWCI: 1.097
Cites: 9
Cited by: 2
Related to: