Midfoot “H-Plate” for operative fixation of posterior wall acetabular fractures: a technical trick and case series
Year: 2025
Type: article
Study design: Observational Study
Abstract: BACKGROUND: Acetabular fractures involving the posterior wall remain challenging to treat, with no gold standard fixation method established. We describe a novel technique using Synthes 2.7 mm non-locking plates originally designed for the midfoot. METHODS: Retrospective review of 120 patients with posterior wall acetabular fractures treated with H-plates at a Level-1 trauma center (2012-2021). Patient demographics, operative parameters, and outcomes were analyzed. RESULTS: The cohort was predominantly male (67.5%) with motor vehicle collisions as the primary mechanism (89.1%). Forty-seven patients (39.1%) received isolated H-plate fixation. Operative time was significantly shorter with isolated H-plates versus additional hardware (2.25 ± 0.60 vs 2.70 ± 0.91 h; p = 0.003). No patients required reoperation for fixation failure. Total hip arthroplasty was required in 9.2% of patients. CONCLUSION: The H-plate offers a simplified approach to posterior wall fixation with promising preliminary results, though longer follow-up is needed.
Authors: C. Julian Clark, Johnathan W. Riley, Alice McCraney, Isaac J. Spears, Drew P. Melancon, Tyler McGee, S. Austin Childress, Peter N. Mittwede, John Morellato, Patrick F. Bergin, Dillon C. O’Neill, Bradley A. Foulke, Matthew L. Graves, George V. Russell
Language: English
Subfield: Surgery
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $2,690
OpenAlex ID: W4416401058