Abstract: A 10-year-old, 7.0 kg, female spayed mixed-breed dog presented for a second opinion regarding lobectomy for a lung mass. Mechanical ventilation was initiated due to hypercapnia (ETCO2 of 61 mmHg) with PIP of 8 cmH2O. The PIP was gradually increased to 16 cmH2O. Six minutes after increasing PIP to 16 cmH2O, ETCO2 suddenly dropped to 0 mmHg and the capnography waveform was lost. The endotracheal (ET) tube connection to the ventilator circuit was intact, and no visible abnormalities were noted on the ET tube. Although the ETCO2 sampling line and water trap, and the monitor were replaced, the ETCO2 remained at 0 mmHg. Meanwhile, suspecting obstruction within the ET tube or the airway, extubation was performed. A large blood clot in the ET tube was noticed. Reintubation was performed using a new ET tube and mechanical ventilation was restarted with PIP set to 10 cmH2O. The ETCO2 showed a reading of 55 mmHg. Five minutes after reintubation, the ETCO2 waveform disappeared again. The inside of ET tube was suctioned, and a blood clot was again identified. Clinicians should be aware of these risks when performing lung lobectomy without one lung ventilation.