Smaller defibrillation lead diameter is associated with less tricuspid regurgitation: a computational modeling study
Year: 2025
Type: article
Abstract: BACKGROUND: Tricuspid regurgitation (TR) adversely impacts mortality and hospitalizations. Transvenous pacemaker and defibrillator leads are associated with worsening TR, underscoring the need for lead designs that minimize adverse tricuspid valve interference. METHOD: We developed a physics-based computational model to evaluate the general effect of lead diameter and material stiffness on the regurgitant orifice area (ROA). Additionally, we simulated three commercial transvenous leads, with diameters of 4.7 Fr, 6.8 Fr and 8.6 Fr, and evaluated the induced ROA. Three pre-implant baseline conditions were considered in all cases: no TR, mild TR and moderate TR. RESULTS: Lead diameter exhibited a predominant effect on ROA compared to material properties. In mild and moderate TR baselines, thicker leads resulted in exponential increases in ROA. The 4.7 Fr lead remained significantly below clinically relevant increases in TR. In contrast, the 6.8 Fr and 8.6 Fr leads induced a transition from mild to moderate TR. CONCLUSIONS: Simulations support the mechanistic expectation that thinner defibrillation leads reduce mechanical interference with tricuspid leaflet coaptation, with potential implications on clinical outcomes.
Source: Expert Review of Medical Devices
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Funder: Medtronic