First evaluation of color K-edge image quality using spectral photon-counting CT combined with two contrast agents: A phantom study

Year: 2025
Type: article
Abstract: PURPOSE: The purpose of this study was to assess the image quality of color K-edge imaging obtained with a spectral photon-counting CT (SPCCT) scanner using a spectral phantom with a mixture of iodine-based and gadolinium-based contrast agents. MATERIALS AND METHODS: A clinical SPCCT scanner prototype was used to scan a spectral phantom. Three dedicated cavities were filled with three contrast agents including iodine alone, gadolinium alone and a mixture of both. Two concentrations of 0.5 and 2 mg/mL were evaluated using nine helical PCCT scans at 120 kVp and 150 mAs. Conventional, color iodine and color K-edge gadolinium images were obtained through a material decomposition algorithm using three basis materials (water, iodine, gadolinium). Attenuation (in Hounsfield unit [HU]), iodine and gadolinium concentrations and task-based transfer function (TTF) were measured on each cavity and image. The noise power spectrum (NPS) was calculated on the phantom's background. RESULTS: ) at 0.5 mg/mL. The value of noise magnitude was 24.75 HU, 0.06 mg/mL and 0.03 mg/mL for conventional, color iodine and color K-edge gadolinium images, respectively. CONCLUSION: Color K-edge imaging helps distinguish between contrast agents while being associated with lownoise magnitude, high-frequency spatial noise and high spatial resolution.