The Impact of Prosthesis-Patient Mismatch on Bioprosthetic Valve Durability After Transcatheter Aortic Valve Replacement
Year: 2025
Type: article
Study design: Observational Study
Abstract: Background: Prosthesis-patient mismatch (PPM) is associated with structural valve deterioration after surgical aortic valve replacement. Limited evidence exists regarding its impact on hemodynamic valve deterioration (HVD) and bioprosthetic valve failure (BVF) following transcatheter aortic valve replacement (TAVR). Methods: Patients who underwent TAVR from 2013 to 2022 at UC San Diego Health with echocardiographic follow-up were retrospectively analyzed. PPM, determined by measured indexed effective orifice area, and outcomes of moderate or severe HVD and BVF were defined per Valve Academic Research Consortium 3 guidelines. Fine and Gray regression, with all-cause death as a competing risk, was used to compare the cumulative incidence of outcomes in patients with and without PPM. Results: = 0.03). Conclusions: In this single-center cohort, PPM was associated with an over two-fold increased risk of moderate or severe HVD and an over four-fold increased risk of BVF following TAVR, emphasizing the importance of pre-implant prevention.
Source: Structural Heart
Authors: Revathy Sampath-Kumar, Erika Padilla, David Torres Barba, Niki Aramburo, Ori Ben-Yehuda, Ehtisham Mahmud
Institution: University of California San Diego
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $2,340
OpenAlex ID: W4417274906