Work

On the speckle‐free nature of photoacoustic tomography

Year: 2009
Type: article
Abstract: PURPOSE: A long-standing conundrum is why photoacoustic tomography (PAT) possesses the unique ability to produce images devoid of speckle artifacts while all other coherent imaging technologies do not. METHODS: In this paper, we explain the inherent mechanism that suppresses speckle in PAT, and the analysis was validated by simulations based on an experimental PAT system. RESULTS: We found that the speckle-free feature of PAT results directly from the optical absorption contrast. CONCLUSIONS: All optical absorbers expand on laser excitation, and therefore all initial photoacoustic pressure rises are positive, which engenders strong correlations among the photoacoustic waves from the absorbers. As a result, prominent boundaries always build up in photoacoustic images and suppress the interior speckle.
Authors: Zijian Guo, Li Li, Lihong V. Wang
Language: English
Open Access status: green
OpenAlex ID: W50584987
FWCI: 2.812
Cites: 18
Cited by: 148
Related to: 10