Work

Design of 3D magnetic sensor array for medical applications

Year: 2010
Type: dissertation
Abstract: When treating tumours with radiotherapy an exact positioning of the organ in question is important in order to minimize damage to adjoining tissue. This thesis was performed at Micropos Medical, a med-tech company situated in Göteborg that has developed an electromagnetic positioning system for real-time positioning inside the human body. The goal of this project was to test and evaluate an industrially manufactured 3D-antenna. Studies have been made to determine if the 3D-antenna can be used to replace the hand wired antenna used today in the system's antenna array and further to evaluate if the 3D-antenna can be used to increase the measurement volume. Different geometries of the 3D-antennas have been tested to collect data and determine precision of the system. The tests show that the primary coil of the 3D-antenna can be used with a mean error radius of 0.63 mm and a standard deviation of 0.47 mm. For 2000 evaluated points 98% of the tested points lie within a 2 mm error margin which is acceptable according to the requirements set by Micropos Medical. The 3D-antennas secondary coils can be used to acquire position, however further testing and development of the algorithm will be needed before it might be possible to use these coils to increase the measurement volume.
Author: Anders Torstensson
Language: English
Open Access status: green
OpenAlex ID: W53207227
Cites: 4
Cited by:
Related to: 10