A calibration device for long-path sensors of atmospheric pollutants. EUR 5472 e. Environment and quality of life
Year: 1976
Type: report
Abstract: The calibration of long-path optical sensors of atmospheric pollutants like the Barringer correlation spectrometer is obtained by a device in which the optical pathlength is varied. By using the device here described the calibration procedure requires about 6 min.for the full length expansion of the cell equivalent to concentration-pathlength variation of 1 : 250. The device is simple and can be installed on an optical bench i.e. for simulation of polluted atmosphere.
Authors: S. Sandroni, Elia Brambilla, M. Ceresoli