Measurement of RF surface efficiency at cryogenic temperatures using a resonant cavity

Year: 1990
Type: conference-paper
Abstract: Exploiting the potential efficiency gain of a normal conducting rf accelerator operated at cryogenic temperatures requires careful preparation of the rf conducting surface. Experimental apparatus has been assembled to study the surface conductivity to rf currents at 425 MHz and 850 MHz through a temperature range from room temperature to 14 K. The apparatus is built around an open-ended coaxial cavity with the cavity tubular ends below the cutoff frequency at resonance. The center conductor in the coaxial cavity is the test sample, and the use of a dielectric stand-off for the center conductor precludes the need for an rf contact joint and facilitates sample changes. The rf testing is conducted under vacuum with low-power rf. A CTI-Cryogenics cryopump coldhead is used for cryogenic temperature cycling of the test cavity. A detailed description of the apparatus and measurement procedures are presented.