Application of Diamond Detectors in Tracking of Heavy Ion Slowed Down Radioactive Beams
Year: 2007
Type: article
Abstract: Results of irradiation of thin Chemical Vapor Deposition (CVD) diamond detectors with low energy: p,α and 7Li beams are presented. Energy resolution: ΔE/E<1% of a single crystal detector was achieved. A coincident measurement with two diamond detectors showed time resolution of 100 ps and efficiency above 70%. Despite a high beam flux reaching 109 particles/s cm2 the tested detectors showed low dead-time and satisfactory radiation hardness. Perspectives of applying thin CVD diamond detectors in monitoring of a slowed down radioactive beam (RIB) are discussed.
Source: AcPPB
Language: English
Subfield: Nuclear and High Energy Physics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W68935361