Abstract: A shipborne detection device has been developed for the detection of heavy metal ions in offshore seawater, utilizing a mercury film electrode. The automatic determination of Cd2⁺ and Pb2⁺ ions in seawater can be conducted without any manual intervention. Waste solutions are collected in a recycling bag for subsequent treatment, thereby preventing environmental pollution. The accuracy, as well as the stability of the measurements, were thoroughly examined. The relative standard deviation for Pb2⁺ was determined to be 7.02% after conducting 10 repeated tests using a standard solution concentration of 5 μg L⁻1 and the relative standard deviation of Cd2+ is 2.65%. Linear calibration curves have been established along with detection limits of 0.31 μg L⁻1 for both Cd2⁺ and Pb2⁺ ions. Ultimately, this shipborne detection device was successfully employed for the in-situ determination of Cd2⁺ and Pb2⁺ ions in seawater.
Source: Figshare
Authors: Lai, Zhiwei, Zhong, Huiling
Language: English
Field: Chemistry
Open Access status: green
OpenAlex ID: W6902211380