Efficient and Cost-Effective Fabrication of PMMA Micropores for Resistive Pulse Sensing
Year: 2024
Type: article
Abstract: This paper delves into the optimization of Poly(methyl methacrylate) (PMMA) micropore size using a comprehensive experimental approach. It begins by highlighting the advantages of PMMA, introducing the concept of drilling a microscale pore in PMMA, and subsequently reducing its size through the application of pressure and heat. The primary goal is to comprehend the influence of pressure, heat, and time on pore dynamics and size alteration. Employing the Taguchi Design of Experiments (DOE), the study systematically investigates the impact of these factors by subjecting a primary 500μm pore to variations in pressure, heat, and time. The results demonstrate the substantial impact of temperature on pore size reduction at higher temperatures, with pressure exhibiting a significant effect up to 1 bar, diminishing thereafter, and time gradually decreasing over the experimental duration. Further exploration focuses on achieving the smallest possible pore size using a 200μm drilled micropore, refining it to below 20μm. Ultimately, the micro-sized pore is utilized in a resistive pulse sensing (RPS) setup, revealing resistance variations associated with different pore sizes. This promising technique holds potential for future applications in RPS setups, facilitating the enhanced analysis and detection of proteins or other minute particles.
Source: Figshare
Author: Rahimi, Parham
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W6902261608
FWCI:
Cites:
Cited by:
Related to: