The Measurement of an NMR Signal with Halbach Arrays

Year: 2019
Type: dissertation
Abstract: The generally high costs of the in vivo imaging technique Magnetic Resonance Imaging (MRI) leads to a low availability of the technology in developing countries. However, to reduce the costs of MRI, a novel approach can be used where permanent magnets in Halbach arrays replace the superconducting magnet normally used in MRI. In this research, a magnet consisting of 4 Halbach rings is constructed to create a magnetic field of 59 mT corresponding to a Larmor frequency of 2.5 MHz for protons. With this new setup, spin echo measurements are performed where a signal-to-noise ratio of 70 is achieved. After that, spin echo trains were successfully measured. The setup can be used to work towards the creation of 2D pediatric brain images with an affordable, portable MRI system that will not exceed the costs of 50,000 US dollars.
Author: Breimer, W
Institution: Leiden University
Language: English
Field: Medicine