Identification and quantification of microplastics in a wide range of aqueous matrices using pyrolysis-gas chromatography-mass spectrometry: method’s performance, greenness evaluation, and analysis of real samples

Year: 2025
Type: article
Abstract: • Eight types of microplastics were analysed in aqueous matrices of varying complexity • Low- and high-complexity samples (e.g., drinking water and wastewater) were analysed • Microwave were used for the first time to digest wastewater and recover microplastics • Pyrolysis-gas chromatography-mass spectrometry was used to analyse microplastics. • The greenness of methods for microplastic analysis was evaluated for the first time This study describes the analysis of microplastics in aqueous matrices of varying complexity using pyrolysis coupled with gas chromatography-mass spectrometry (Py-GC-MS). Depending on the filtration rate of the analysed matrix, sample treatment prior to Py-GC-MS consisted of filtration on 0.7 µm glass fibre membranes (low-complexity matrices), or microwave-assisted digestion of samples with hydrogen peroxide, followed by filtration as above-mentioned (high-complexity matrices). Low-complexity matrices included demineralised water (DMW), mineral water in polyethylene terephthalate (PET) bottles (BMW), drinking water from a public aqueduct (DW), and DW further treated with activated carbon and distributed in public fountains (FOW). High-complexity matrices included effluent wastewater from primary sedimentation (WW-PS), biological oxidation (WW-BO), clariflocculation (WW-CL), and quaternary treatment (WW-QT). Sample digestion significantly accelerated the filtration process of high-complexity matrices. In most cases, absolute value of matrix effect was <40%, except for PET in WW-PS (-54%) and WW-CL (-56%), and polystyrene in WW-PS (75%). Apparent recovery, studied in DMW, DW, and WW-BO matrices, was 90±24, 87±25, and 67±19, respectively. The methods were evaluated for their greenness, achieving scores higher than or comparable to those of previously published Py-GC-MS procedures. Poly(methyl methacrylate) and PET were never detected, while different levels of contamination were found for polystyrene, polyethylene, polypropylene, polycarbonate, polyamide, and polyvinyl chloride, depending on the sample type. The lowest contaminations were determined in BMW and DMW. FOW was more contaminated than DW (five microplastics detected at 2.7-138 µg/L vs. two microplastics detected at <0.62-23 µg/L) and concentrations in WWs were in the order WW-QT<WW-CL<WW-PS<
Language: English
Subfield: Pollution
Open Access status: gold
Estimated APC paid: $1,550
OpenAlex ID: W7084035437
FWCI: 0.67
Cites: 39
Cited by: 3
Related to: