Work

Estimation of Physiological Tremor from Accelerometers for Real-Time Applications

Year: 2011
Type: article
Abstract: Accurate filtering of physiological tremor is extremely important in robotics assisted surgical instruments and procedures. This paper focuses on developing single stage robust algorithms for accurate tremor filtering with accelerometers for real-time applications. Existing methods rely on estimating the tremor under the assumption that it has a single dominant frequency. Our time-frequency analysis on physiological tremor data revealed that tremor contains multiple dominant frequencies over the entire duration rather than a single dominant frequency. In this paper, the existing methods for tremor filtering are reviewed and two improved algorithms are presented. A comparative study is conducted on all the estimation methods with tremor data from microsurgeons and novice subjects under different conditions. Our results showed that the new improved algorithms performed better than the existing algorithms for tremor estimation. A procedure to separate the intended motion/drift from the tremor component is formulated.
Source: Sensors
Language: English
Open Access status: gold
Estimated APC paid: $3,137
OpenAlex ID: W2071890760
FWCI: 4.08
Cites: 36
Cited by: 128
Related to: 8