Detection of Carbapenemases by Real-Time PCR and Melt Curve Analysis on the BD Max System
Year: 2014
Type: article
Abstract: We developed a multiplex SYBR green real-time PCR for the BD Max instrument (BD Diagnostics, Sparks, MD) to detect a panel of carbapenemases. The assay was evaluated with 152 consecutive isolates sent to the German National Reference Laboratory, and 65/65 of the carbapenemase-positive and 87/87 of the carbapenemase-negative strains were identified correctly.
Source: Journal of Clinical Microbiology
DOI: 10.1128/jcm.00373-14
Institutions: University Hospital Heidelberg, Ruhr University Bochum
Language: English
Subfield: Microbiology (medical)
Field: Medicine
Domain: Health Sciences
Open Access status: green
Funder: Robert Koch Institut
OpenAlex ID: W2109741674