Work

Irreversible intracardiac defibrillator circuit damage induced by pulsed-field ablation: Mechanistic insights

Year: 2026
Type: article
Abstract: Pulsed-field ablation (PFA) is a non-thermal modality that induces cardiomyocyte death via irreversible electroporation, achieving effective pulmonary vein isolation (PVI) while sparing surrounding tissues.1-2 Although PFA is generally considered safe in patients with cardiac implantable electronic devices (CIEDs),3-5 its strong electric fields may interfere with device function, potentially causing circuit damage or inappropriate responses.6 Recently, irreversible internal damage to an implantable cardioverter-defibrillator (ICD) was reported when PFA was delivered directly to the superior vena cava (SVC) coil of the ICD lead.
Language: English
Field: Medicine
Open Access status: gold
Estimated APC paid: $1,100
OpenAlex ID: W7119025085
FWCI: 14.47
Cites: 6
Cited by: 3
Related to: