Análise da resistência mecânica e fractográfica de placas de titânio do sistema 1,5mm e dois métodos de fixação em fraturas simuladas em corpo de hemimandíbula de alumínio

Year: 2013
Type: dissertation
Abstract: The purpose of this study was to analyze the mechanical strength of two brands of 1.5 mm system fixation through mechanical linear load test in two fixation methods in aluminum hemimandibles with simulated body fractures and perform fractographic analysis of the samples.Four aluminum structures similar to hemimandibles were fabricated with a linear sectioning simulating a fracture of mandibular body and two hemimandibles were used in each group for posterior fixation with plates and screws with two national brands: Toride® and Traumec®.The group I was fixed with a straight titanium plate with four holes in the compression zone and another one in the tension zone.Group II was fixed with a straight plate with four holes in the neutral zone.For mechanical testing, a universal mechanical testing Instron ® Model 4411 was used and the test was made until final failure of the system (fracture plate).Then, the samples were submitted to fractographic analysis through a scanning electron microscope FEI ®.A analysis of variance (ANOVA) Two-way and Tukey test with 5% significance ( = .05)were applied for statistical analysis.The fixation method with two straight plates showed a higher load resistance (p<0.01) when compared to the group of only one plate, regardless of the brand.The brand Toride® showed an average of peak displacement higher than Traumec® when the groups with only one plate with the extension of 5mm were evaluated.All the samples fractured by ductile overload, characterized by alveolar micromechanisms.
Language: Portuguese
Subfield: Surgery
Field: Medicine
Open Access status: gold
OpenAlex ID: W2222039779
Cites: 58
Cited by:
Related to: 8