Work

Ion selective phosphotungestate and beta-cyclodextrin based membrane electrodes for stability-indicating determination of midodrine hydrochloride.

Year: 2013
Type: article
Abstract: This paper reports the construction and evaluation of two ion selective electrodes for the determination midodrine hydrochloride (MD) by direct potentiometry in pure drug substance and in tablet formulations. Precipitation based technique was used for fabrication of the first membrane sensor (sensor 1) using phosphotungestate (PT) and dioctylphthalate (DOP) as cation exchanger and solvent mediator, respectively. beta-cyclodextrin (beta-CD)-based technique with PT as a fixed anionic site in PVC matrix was used for fabrication of the second membrane sensor (sensor 2). The proposed sensors showed fast, stable Nernstian responses of 54 and 56 mV/decade for sensors 1 and 2, respectively, across a relatively wide MD concentration range (1x 10(-4) to 1 x 10(-1) mol/L and 5 x 10(-5) to 1 x 10(-1) mol/L for sensor 1 and 2, respectively) in the pH range of 5-7. Sensor I and sensor 2 can be used for three and two weeks, respectively without any measurable change in sensitivity. The suggested electrodes succeeded to determine intact MD in the presence of up to 10% of its degradation product and displayed good selectivity in presence of common inorganic and organic species.
Source: PubMed
Institution: Cairo University
Language: English
Field: Chemistry
Open Access status: green
OpenAlex ID: W2410204091
FWCI: 0.96
Cites: 3
Cited by: 21
Related to: 9