A portable gamma camera for the optimization of the patient dosimetry in radioiodine therapy of thyroid diseases
Year: 2021
Type: conference-paper
Abstract: Molecular radiotherapy is an efficient treatment modality of benign and malign thyroid diseases. However, there is still a need to better assess the dose delivered to target tissues and organs-at-risk in order to optimize for each patient the activity to be administered according to the objectives of disease control (destruction of tumor residues, restoration of thyroid function or hypothyroidism) while maintaining the risk of toxicity at a justifiable level. In that context, our objective is to develop a high-resolution mobile gamma camera specifically designed to accurately measure the radiotracer biokinetics at the patient's bedside during treatment planning and therapeutic dose verification. A first feasibility prototype of the mobile camera with a 5×5 cm 2 field of view was developed for the treatment of benign and malign thyroid diseases with 131 I, leading to promising results. We are currently developing a new prototype for clinical use with extended field of view (10×10 cm 2 ). It consist of a 3D-printed parallel-hole tungsten collimator coupled to a 1 cm thick CeBr3 scintillator, readout by an array of 6×6 mm 2 Silicon Photomultipliers. Preliminary results show a energy resolution of 7.1% and a FWHM spatial resolution around 1 mm at 356 keV. A detailed description of the camera optimisation (collimator and shielding design, intrinsic spatial performance, counting rate capabilities) will be presented.
Source: HAL
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W7124233998