Acute Atrial Pacing Threshold Elevation in an Auricular Leadless Pacemaker During Monopolar Pulsed Field Ablation of the Right Inferior Pulmonary Vein: A First‐in‐Human Case
Year: 2026
Type: article
Study design: Case Report
Abstract: INTRODUCTION: Leadless atrial pacemakers are increasingly used, yet interactions with pulsed field ablation (PFA) remain poorly described. METHODS AND RESULTS: We report the first documented case of transient loss of capture and pacing threshold elevation in an AVEIR leadless pacemaker temporally associated with monopolar PFA during right inferior pulmonary vein ablation. Threshold increased from 0.75 V @ 0.2 ms to 1.25 V @ 0.4 ms, recovering within 24 h. CONCLUSION: Monopolar PFA can transiently interfere with atrial leadless pacing, emphasizing the need for device monitoring during ablation.
DOI: 10.1111/jce.70259
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W7124277338
