Evaluation of the Truenat® chip-based real-time PCR platform for infectious disease diagnostics
Year: 2026
Type: article
Abstract: Objective: - MTB, malaria, leptospirosis), focusing on analytical performance, repeat rates, and time-to-result. Methods: Remnant clinical specimens and commercial controls were analyzed under routine conditions. Performance metrics included accuracy, reproducibility, limit of detection (LOD), linearity, and specificity, when applicable. Each analyte was tested in replicates, and results were compared with established reference methods. Results: = 0.9996) with an observed LOD of 3000 copies/mL. Leptospira assays yielded 100% accuracy with a LOD of 750 copies/mL. Performance generally matched manufacturer specifications, though low-burden MTB samples were less consistently detected. Conclusion: Truenat assays showed reliable analytical performance, especially for COVID-19. While MTB detection remains more robust with Xpert Ultra in low-burden cases, Truenat provides a viable point-of-care alternative in resource-limited settings. For malaria and leptospirosis, broader clinical validation is needed before routine implementation in our service.
Source: Practical Laboratory Medicine
Authors: João Renato Rebello Pinho, Gabriela Rampazzo Cruz, Priscila de Meira Oliveira, Fernanda de Mello Malta, Roberta Cardoso Petroni, Roberta Miraglia, Ana Claudia Campana, Alice Friedenberg de Ulhoa Cintra, Ricardo Andreotti Siqueira, Rubia Anita Ferraz Santana, Fabiane Camargo Gomes Nunes, Marinês Dalla Valle Martino, André Mario Doi, Nair Hideko Muto, Cristóvão Luis Pitangueira Mangueira
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $2,140
OpenAlex ID: W7128020119