Fracture and Perforation of Coronary Sinus Lead due to Retained Guide Wire During Index Cardiac Resynchronization Therapy Implantation
Year: 2026
Type: article
Abstract: Our case reinforces that the retained guidewire technique should be abandoned for coronary sinus lead stabilization during cardiac resynchronization therapy. Although rarely encountered today, patients previously implanted using historical methods may present with severe mechanical complications and high-risk extraction profiles. Early recognition and referral to experienced lead extraction centers is essential.
Source: Journal of Arrhythmia
DOI: 10.1002/joa3.70293
Institution: Hacettepe University
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $1,540