Abstract: Objective This study assessing the performance of five intraocular lens (IOL) calculation methodologies in cataract patients with previous radial keratotomy (RK) history.Methods Five formulas predicted refractive outcomes, with prediction errors (PE) calculated by comparing formula-derived spherical equivalents to manifest refraction. Additional statistical analysis examined correlations between formula-specific PE and biometric parameters including axial length (AL) and corneal curvature.Results Overall, Barrett True-K and Hoffer QST showed minimal prediction bias (PE: −0.06D and −0.07D; p > 0.05). PEARL-DGS had the best stability (SD: 0.97) and led in ±0.25D accuracy (20.8%, p = 0.665). In AL subgroups, Barrett True-K showed superior accuracy for medium-long eyes (26.0–30.0 mm; PE closest to zero), while EVO 2.0 yielded significant hyperopic error in short AL eyes (<26.0 mm). Significant negative correlations between flat keratometry and PE were identified for the Hoffer QST, Kane, and Pearl DGS formulas (all p < 0.01). In eyes with flat keratometry < 39.0 D, EVO 2.0, Kane, and Pearl DGS exhibited statistically significant hyperopic biases, while Barrett True-K and Hoffer QST maintained accuracy without systematic error.Conclusion The results indicate that for post-RK patients, the Barrett True-K formula exhibits enhanced accuracy in IOL power calculation. Newer AI-based formulas show potential but currently remain limited by a systematic hyperopic bias in this population.
Source: Journal of Investigative Surgery
Institutions: Beijing Tongren Hospital, Capital Medical University
Language: English
Subfield: Ophthalmology
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $2,970
OpenAlex ID: W7130319071