Work

Accuracy and performance of the Yuwell Anytime CT3 continuous glucose monitoring device

Year: 2026
Type: article
Abstract: BACKGROUND: There have been many recent advances in diabetes technology, and continuous glucose monitoring (CGM) systems have grown in use and popularity. AIMS: This study evaluated the accuracy and efficacy of the Yuwell Anytime CT3 CGM system in participants with type 2 diabetes. METHODS: Forty participants were recruited to take part. They were asked to monitor their glucose levels for 10 days using two devices, a continuous glucose sensor (Yuwell Anytime) and an Accu-chek blood glucose finger-prick test. FINDINGS: <0.0001). The reliability of the sensor was preserved across the full observed range of glucose concentrations. CONCLUSIONS: The Yuwell Anytime had excellent agreement with a standard finger-prick blood test of glucose in free-living type-2 diabetics.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W7130330809
FWCI:
Cites: 27
Cited by:
Related to: