Failure Mode and Effects Analysis in Reducing the Incidence of Sterilization Cycle Interruptions in Low-Temperature Hydrogen Peroxide Plasma Sterilizers

Year: 2026
Type: article
Abstract: Liangying Yi,1,2,* Ling You,3,* Wei Pan,1,2 Kaili Wu,2,4 Hongxia Zhang1,2 1Department of Sterile Processing Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, People’s Republic of China; 2Key Laboratory of Birth Defects and Related Diseases of Women and Children (Sichuan University), Ministry of Education, Chengdu, Sichuan, People’s Republic of China; 3Lung Cancer Center/Lung Cancer Institute, West China Hospital, Sichuan University/West China School of Nursing, Sichuan University, Chengdu, Sichuan, People’s Republic of China; 4Department of Pediatric Hematology and Oncology Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, People’s Republic of China*These authors contributed equally to this workCorrespondence: Wei Pan, Email [email protected] Kaili Wu, Email [email protected]: This study aimed to investigate the efficacy of failure mode and effects analysis (FMEA) in optimizing the sterilization process for low-temperature hydrogen peroxide plasma sterilizers.Methods: Purposive sampling was used to select low-temperature hydrogen peroxide plasma sterilizers from a hospital in China as the study objects. The operational data for 1551 sterilization cycles in the low-temperature hydrogen peroxide plasma sterilizers, from May 2024 to November 2024, were classified into the control group. The operational data for 1492 sterilization cycles in the low-temperature hydrogen peroxide plasma sterilizers, from January 2025 to June 2025 following the implementation of the FMEA-based risk management scheme, were classified into the experimental group. The incidence of sterilization cycle interruptions, the risk priority number scores for high-risk failure modes, and the incidence of high-risk failure modes in both groups were investigated.Results: After the optimization of the operating procedures, the incidence of sterilization cycle interruptions decreased from 5.48% (85/1551) to 0.87% (13/1492); the risk priority number scores and the incidence of high-risk failure modes were also decreased (P < 0.05).Conclusion: The FMEA-based risk management scheme could optimize the sterilization process for low-temperature hydrogen peroxide plasma sterilizers and ensure the quality of medical device sterilization.Keywords: failure mode and effects analysis, low-temperature hydrogen peroxide plasma sterilizers, sterilization interruptions, sterilization quality, risk management
Language: English
Subfield: Microbiology
Open Access status: green
OpenAlex ID: W7130497152
FWCI:
Cites:
Cited by:
Related to: