Transforming the Design Process of Orthoses: Evaluating the Accuracy and Reliability of a 3D Scanning Process for Pediatric Ankle-Foot Orthoses Using Digital Technology
Year: 2024
Type: dissertation
Abstract: A major challenge to adopting digital workflows for the design and fabrication of ankle-foot orthoses is the difficulty in controlling lower-limb positioning and alignment during 3D scanning. This study is the first to evaluate the accuracy and reliability of a 3D scanning process for pediatric AFOs by comparing direct lower limb scans to cast models. Participants were recruited in collaboration with the Orthotics and Prosthetics Department at Holland Bloorview Kids Rehabilitation Hospital. Through a data-driven and quantitative approach, the findings confirm the feasibility of using 3D scanning and CAD for AFO design and fabrication. The study quantitatively measured sagittal plane alignment and analyzed shape differences between direct scans and cast models using digital tools. Results indicated that direct scans closely matched the cast models, with variability largely attributed to limb positioning, alignment, and compression during the scanning process. These findings will inform future developments in digital AFO design and fabrication.
Source: TSpace (University of Toronto)
Author: Matton, Connor Douglas Matthew
Institution: University of Toronto
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W7133030528