Work

Non-Contact Heart Rate Estimation via Higher Harmonic Analysis Using 24-GHz Doppler Radar: Validation in Humans and Anesthetized Cat

Year: 2026
Type: article
Abstract: This study presents a harmonic-based method for non-contact heart rate (HR) estimation from continuous-wave (CW) Doppler radar signals, validated across multiple species including humans and small animals (cat). Traditional frequency-domain methods struggle when the HR fundamental frequency is weak or overlaps with respiratory components. The proposed approach addresses this by identifying three higher-order HR harmonics (2nd, 3rd, and 4th) then reconstructing the HR fundamental frequency from their integer ratios (3/2, 4/3, 2/1). The algorithm processes 20-s sliding windows (1-s overlap) using bandpass filtering to remove respiratory components and HR fundamental while preserving higher harmonics, followed by Power Spectral Density (PSD) analysis. When a complete harmonic set cannot be found, the proposed algorithm switches to harmonic pair detection, enhancing robustness when one harmonic is absent or attenuated. Besides, an adaptive tolerance mechanism enables detection under non-ideal conditions. The method was validated using a public human dataset and an experimental cat dataset with varied positions (supine/prone) and anesthesia levels (1–3% isoflurane). For humans, the algorithm achieved HR Accuracy consistently above 98% with an average RMSE of 1.33 bpm (MAPE: 1.29%, MAE: 0.86 bpm) and Bland-Altman bias below 0.9 bpm. For the cat dataset, performance was even better with HR Accuracy remaining above 99%, an average RMSE of 0.39 bpm (MAPE: 0.22%, MAE: 0.30 bpm), and bias below 0.14 bpm.
Source: Signals
Language: English
Open Access status: gold
Estimated APC paid: $1,207
OpenAlex ID: W7133550233
FWCI:
Cites: 44
Cited by:
Related to: